A THREE-PHASE INDUCTIVE SENSOR FOR IN VIVO MEASUREMENT OF ELECTRICAL ANISOTROPY OF BIOLOGICAL TISSUES

Authors

  • Yukio Kosugi Tokyo Institute of Technology
  • Tadashi Takemae Shizuoka University
  • Hiroki Takeshima Shizuoka University
  • Atsushi Kudo Tokyo Medical and Dental University
  • Kazuyuki Kojima Tokyo Medical and Dental University
  • Naofumi Tanaka Tokyo Medical and Dental University

DOI:

https://doi.org/10.7251/COMEN1401001K

Abstract

Biological tissue will have anisotropy in electrical conductivity, due to the orientation of muscular fibers or neural axons as well as the distribution of large size blood vessels. Thus, the in vivo measurement of electrical conductivity anisotropy can be used to detect deep-seated vessels in large organs such as the liver during surgeries. For diagnostic applications, decrease of anisotropy may indicate the existence of cancer in anisotropic tissues such as the white matter of the brain or the mammary gland in the breast.

In this paper, we will introduce a new tri-phase induction method to drive rotating high-frequency electrical current in the tissue for the measurement of electrical conductivity anisotropy. In the measurement, three electromagnets are symmetrically placed on the tissue surface and driven by high-frequency alternative currents of  0 kHz, modulated  with
1 kHz 3-phase signals. In the center area of three magnets, magnetic fields are superimposed to produce a rotating induction current. This current produces electrical potentials among circularly arranged electrodes to be used to find the conductivity in each direction determined by the electrode pairs. To find the horizontal and vertical signal components, the measured potentials are amplified by a 2ch lock-in amplifier phase-locked with the 1 kHz reference signal. The superimposed current in the tissue was typically 45 micro Amperes when we applied 150 micro Tesla of magnetic field. We showed the validity of our method by conducting in vitro measurements with respect to artificially formed anisotropic materials and preliminary in vivo measurements on the pig’s liver.

Compared to diffusion tensor MRI method, our anisotropy sensor is compact and advantageous for use during surgical operations because our method does not require strong magnetic field that may disturb ongoing surgical operations.

References

[1] P. Steendijik, G. Mur, T. E. Van der Velde, J. Baan, The Four-Electrode Resistivity Technique in Anisotropic Media: Theoretical Analysis and Application on Myocardial Tissue in Vivo, IEEE Trans. Biomed Eng., Vol. 40−11 (1993) 1138−1148.

[2] Y. Hiramatsu, J. W. Buchanan, S. B. Knisley, L. S. Getters, Rate Dependent Effects of Hypox-ia on Internal Longitudinal Resistance in Guinea Pig Papillary Muscles, Circ Res., Vol. 63−5 (1988) 923−929.

[3] D. S. Holder (ed.), Electrical Impedance Tomography: Methods, History and Applications, IOP Publishing (2005).

[4] K. Pek-Lan, H. T. Lucullus Leung, et al., White Matter Anisotropy in Post-Treatment Childhood Cancer Survivors: Preliminary Evidence of Association With Neurocognitive Function, Journal of Clinical Oncology, Vol. 24−6 (2006) 884− 890.

[5] E. Furman-Haran, E. Eyal, M. Shapiro-Feinberg, D. Grobgeld, T. Golan, et. al., Breast Cancer Detection and Diagnosis based on Diffusion Tensor Imaging, Proc Intl. Soc. Mag. Reson. Med., Vol. 19 (2011) 515.

Downloads

Published

2014-09-24